Overview

Basic Info.

Model NO.
YJ-D2000
Screen
LCD Touch Screen
Input Power
1500W
Water Temperature
5degree to 30 Degree
Water Input Pressure
0.15~0.6MPa
Pressure and Pulse
Auto Measuring
Market
Global
Color
White
Package Size
680*550*1650mm
Transport Package
Wooden Package
Specification
380*400*1380mm(L*W*H)
Trademark
Foremed
Origin
China
HS Code
90189040
Production Capacity
2000sets Per Month

Product Description

Product Description

YJ-D2000 Hemodialysis Machine



This haemodialysis machine is specially designed, produced and sold for renal failure patients to receive haemodialysis.It is applicable for chronic renal failure and other blood Purification treatment.

Choosing Reasons:
1.Classical and perfect model designed.
2.Two persons can be trained freely for first time.
3.All accessories or spare parts with tools can be along with hemodialysis machine can be set along.
4.Rich knowledge engineer and sales can be serve for you.
5.One year guarantee time,5years spare parts supplied at least.
Application Purpose of this Device
YJ-D2000 haemodialysis machine is applicable for chronic renal failure and other blood Purification treatment.
This device should be used in medical units.
This device is specially designed, produced and sold for renal failure patients to receive haemodialysis, which is not allowed to be used for other purposes.

Forms of Therapy: 
Haemodialysis, Isolated Ultrafiltration, Sequential Ultrafiltration, Hemoperfusion, etc.

Main Features:
 Intelligent Double Operation System 
 LCD touch screen with button interface 
 Emergency power 30Mins (Optional) 
 Blood Pump 
 Spare Pump (for standby and also can be used for hemoperfustion).
 Heparin Pump
 Hydraulic compartment(Balance chamber + UF pump).
 Operation, Alarm information Memory function.
 A/B Ceramic proportion pump , High precision, Corrosion-proof, Accuracy.

Dialysate:
Dialysate temperature: preset range 34.0ºC~39.0ºC
Dialysate flux: 300~800 ml/min
Dialysate concentration: 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm
Dialysate mixing ratio: can set variety ratio.
UF rate Flow range: 0 ml/h ~4000 ml/h
Resolution ratio: 1ml
Precision: ±30 ml/h
Extracorporeal Part:
Venous pressure: -180 mmHg ~+600 mmHg, ±10 mmHg
Arterial pressure : -380 mmHg ~+400 mmHg, ±10 mmHg
TMP  pressure : -180 mmHg ~+600 mmHg, ±20 mmHg                           
Blood pump flow range: 20 ml/min ~400 ml/min(diameter:Ф6 mm)
Spare pump flow range: 30 ml/min ~600 ml/min(diameter:Ф8 mm)
Resolution ratio: 1 ml
Precision: error range ±10ml or 10% of reading
Heparin Pump:
Syringe size: 20, 30, 50 ml
Flow range: 0 ml/h ~10 ml/h
Resolution ratio: 0.1ml
Precision: ±5%

Sanitize :
1.Hot decalcification :
Time: about 20 minutes
Temperature: 30~60ºC, 500ml/min.
2.Chemical disinfection :
Time: about 45 minutes
Temperature: 30~40ºC, 500ml/min.
3. Heat disinfection :
Time: about 60 minutes
Temperature: >85ºC, 300ml/min.                                                
Storage Environment  Storage temperature should be between 5ºC~40ºC, at relative humidity of no more than 80%.

Monitoring System :
Dialysate temperature: preset range 34.0ºC~39.0ºC, ±0.5ºC
Blood leak detection: Photochromic
Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate
Bubble detection: ultrasonic
Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow
Conductivity: acoustic-optic, ±0.5%

Technical specifications

Main Parameters For Hemodialysis Machine:                                                   

Model YJ-D2000
Power  1500W
Working environment  temperature 10ºC ~30ºC (at relative humidity of no more than 70%.)
Water input pressure  0.15 MPa   ~0.6 MPa
Water input temperature  10ºC~30ºC
Machine Size  380mm×400mm×1380mm (L*W*H)                              
Package Size 680*550*1680mm
Net/Gross Weight  88KG/93kg
 
All disposable accessories or other accessories such as Reprocessor,Dialyzer Rinse Machine,RO Water Machine,Dialysate Dosing Device,Dialysis Chair,Dialyzer,Hemoditoxifier,Anti-cross Infection Joint,Test Paper,Dialysis Catheter,Disposable A.V. Fistula Needel,Disinfectant,Dialysis Concentrate and Powder,Haemodialysis tubing package,etc.
Note:One year quality guarantee and 24 hours phone service for all the products supplied.


Relative Products:





Our Company

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now