| Customization: | Available |
|---|---|
| Customized: | Customized |
| Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency

FHR | Transducer | Multi-crystals, wide wave beam, pulsed-wave working method, high sensitivity |
Strength | <5mW/cm2 | |
Working Frequency | 1.0MHz | |
Signal Processing | Special DSP system and modern recogniti0on | |
Measurement Range | 50-240bpm | |
Alarm Range | High limit:160,170,180,190bpm Low Limit:90,100,110,120bpm Maximum Audio Output Power:1.5 Watt | |
TOCO | Measurement Range | 0-100 Units |
Display | LCD shows the FHR trace, TOCO traced, FM, Doctor Event Mark, Time, Volume etc monitor state, and also it can store and playback. | |
Dimention | 350*320*85mm | |
Net Weight | 3.5kg | |
SpO2 | Measurement Range | 70%-99% |
Measurement Accuracy | ±3% error | |
NIBP | Measurement Range | SYS 6.7~32.0kPa(50~240mmHg) MEAN: 3.4~26.6kPa(25~200mmHg) DIA: 2.0~24.0ka(15~180mmHg) |
NIBP Accuracy | ±1.1kPa(±8mmHg)or ±5% of results, taking the bigger one; | |
HR | HR Measurement Range | 40bpm~240bpm |
Measurement Accuracy | ±2 bpm or ±5% of the results, taking the bigger one | |
Standard | SPO2, MHR, NIBP, TEMP, ECG, RESP, FHR, TOCO, FM | |
Optional | Twin Monitoring, FAS(fetal acoustic simulator) | |