| Customization: | Available |
|---|---|
| Classification: | Imaging Diagnostic Equipment |
| Type: | Blood Gas Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency


Parameter | Accuracy | Range | Precision |
K+ | ±3.0% | 2.0-9.0 mmol/l | ≤1.5% |
Na- | ±3.0% | 100-180mmol/L | ≤1.5% |
CI- | ±3.0% | 65-140 mmol/L | ≤1.5% |
iCa2+ | ≥1.00 mmo/L,±5.0%; <1.00 mmol/L,±0.05mmol/L | 0.25-2.50 mmol/L | ≤1.5% |
pH | ±0.04 | 6.500-8.000 | ≤0.02 |
pCO2 | ≥62.0mmHg,±8.0%; <62.0mmHg,±5.0 mmHg | 10.0-150.0 mmHg | ≥62.0mmHg, CV≤4.0%; <62.0mmHg, SD≤2.5 mmHg |
pO2 | ≥50mmHg,±15.0%; <50mmHg,±7.5 mmHg | 10-700 mmHg | ≤5.0% |
GIu | ≥4.0 mmo/L, ±10.0%; <4.0 mmol/L, ±0.4mmol/L | 1.1-38.0 mmol/L | ≥4.0 mmol/L, CV≤5.0%; <4.0 mmo/L, SD≤0.2 mmol/L |
Lac | ≥5.00 mmol/L,±12.0%; <5.00 mmo/L,±0.6mmol/L | 0.50-20.00 mmol/L | ≥5.00 mmol/L, CV≤6.0%; <5.00 mmol/L, SD≤0.3 mmol/L |
Hct | ≥50%PCV, ±6.0%; <50%PCV, ±3%PCV | 10%-70%PCV | ≥50%PCV, CV≤3%; <50%PCV, SD≤1.5%PCV |