Customization: | Available |
---|---|
Classification: | Nucleic Acid Purification Extraction System |
Type: | Nucleic Acid Purification Extraction System |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Accurate:
Accurate temperature control and rapid temperature rise. Adopt single row lifting mode to reduce liquid evaporation and strictly control the reaction system.Kit name | Specification | Kit name | Specification |
Nucleic acid extraction or purification reagents | 20T/box | Viral DNA/RNA Extraction Kit | 20T/box |
Nucleic acid extraction or purification reagents | 32T/box | Viral DNA/RNA Extraction Kit | 32T/box |
Nucleic acid extraction or purification reagents | 64T/box | Viral DNA/RNA Extraction Kit | 64T/box |
Nucleic acid extraction or purification reagents | 96T/box | Viral DNA/RNA Extraction Kit | 96T/box |
Nucleic acid extraction or purification reagents(not pre-packed) | 50T/box | Viral DNA/RNA extraction kit (not pre-packed) |
50T/box |
screen operation | |
Multi-mode operation | Scanning method, list method,shortcut,viewing method |
Processing volume | 20ul-1000ul |
Sample throughput | 1-48 |
Magnetic bead recovery |
98% |
Orifice Type |
96-well 2.2ml deep-well plate |
Difference between extraction wells |
CV<3% |
Heating Temperature |
12 independent wrap-type heating modules, the pyrolysis and elution temperature can be customized according to requirements (temperature range: 0-99°C), and the temperature sensing accuracy can reach 0.1°C |
Purification Sensitivity |
Positive detection rate of 100 copy samples > 99% |
Kind of reagent |
Magnetic Bead Method Open Platform |
Extraction time |
8-35min |
Internal procedure |
Can store at least 200 experimental program |
Program management |
Powerful program editing function, To meet the needs of different reagents, Can realize the import and export of USB program |
Data storage |
Storage, built-in SD card |
Security door design |
After the safety door is opened, the program operation is automatically suspended, and the program can continue to run after the safety door is closed to avoid harming the human body |
Built-in air duct |
Built-in air duct to avoid cross contamination |
Flashlight |
Long life multi-point LED |
Disinfection/Decontamination Methods |
UV lamp |
Input power |
AC85-264V,47Hz-63Hz,450W |