Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Incubator, High Quality Platelet |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
High quality Medical Cardiac First Aid Portable AED automated external defibrillator calibrator portable defibrillator
Output
|
Biphasic Truncated Exponential
|
Energy Sequence
|
150, 150, 200J
|
Charge Time
|
8 sec. to 150J; 10 sec. to 200J (New Battery)
|
Analysis Time
|
9 sec.
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery
|
Less than 30 seconds |
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks
|
Less than 35 seconds
|
The maximum time from initially switching power on to readiness for discharge
|
Less than 25 seconds
|
Audible Prompts
|
22 audible prompts
|
Visual Prompts
|
LED prompts
|
Controls
|
Two buttons-ON/Off, Shock
|
Output Energy Accuracy
|
±15% into any impedance from 25 to 175Ω
|
The Maximum Voltage
|
1200±50V
|
Output disabled when the patient impedance is outside limits
|
20Ωto 200Ω
|
Category
|
Specification
|
||
Dimensions
|
303 x 216 x 89 mm
|
||
Weight
|
2.0 kg
|
||
Operating Temperature
|
0ºC to 40ºC
|
||
Operating Humidity
|
Relative humidity between 30% and 95% (non-condensing)
|
||
Storage Temperature
|
-20ºC to 55ºC
|
||
Storage Humidity
|
Up to 93% (non-condensing)
|