Customization: | Available |
---|---|
Classification: | Anesthesia Video Laryngoscope |
Type: | Anesthesia Video Laryngoscope |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
YJ-CR-31D Anesthesia video laryngoscope
Video Laryngoscopesolution can be used to provoke patients' glottis and exposeepiglottis,guidemedical staff to accurately intubate airway for anesthesia or first aid, and also for oral inspection and treatment.
Parts
|
Description
|
Technical Index
|
Machine
|
Display
|
3" LCD Full View
|
Resolution Ratio
|
1280*720(RGB)
|
|
Viewing Angle
|
2609
|
|
Power
|
<2W
|
|
Display front and back
left & right rotation angle |
0°~150°0°-270°
|
|
Weight
|
225g
|
|
Camera
|
lluminance
|
£400 LUX
|
Resolution Ratic
|
1280*720 px
|
|
Image /Video
Function |
Image / Video function
|
Yes
|
Output:
|
USB output,automatic overage of full recorc
Easy to storage and establish image/video |
|
Memory type
|
Micro SD card 16GB
|
|
Battery
|
Battery type
|
Rechargeable lithium battery
|
Capacity
|
1500mAh
|
|
Cycle life of battery
|
>500 times
|
|
Battery working time
|
>200min
|
|
Charging Time
|
<3(hrs)
|
|
Charging Port
|
Micro USB
|
|
Power
Adapter |
Input
|
100-250V,50 Hz.
|
Output
|
5V1A
|
The best choice for difficult and routine airways
|
|||
Item Code
|
Type
|
O.D.Size
|
Tube Size
|
CR-VLS-P
|
Pediatric
|
3.9mm
|
4.0-6.0 mm
|
CR-VLS-A
|
Adult
|
5.5mm
|
6.0-10 mm
|
Sterilize Method
|
2.0% Glutaral fully soak for 10 minutes Or
0.5%-0.6% Cidex OPA for more than 5 minutes |
* Reusable blade parameters
|
|||||||
|
|
|
|
|
|
|
|
Blade
|
MIL00
|
MILO
|
MAC1
|
MAC2
|
MAC3
|
MAC4
|
D blade
|
Application
|
Neonate
|
Neonate
|
Pediatric
|
Small Adult
|
Adult
|
Large Adult
|
Difficult airway
|
Intubation
Inner Diameter |
1.5-2.5mm
|
2.5-4.0mm
|
4.0-6.0mm
|
5.0-6.0mm
|
6.0-9.0mm
|
7.5-10.0mm
|
7.5-10.0mm
|
Length(cm)
|
6.0
|
8.2
|
10.45
|
12.11
|
14.25
|
16.08
|
13.27
|
Sterilize
Method |
2.0% Glutaral fully soak for 10 minutes Or 0.5%-0.6% Cidex OPA for
more than 5 minutes |