Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | X Ray Equipment |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Digital Mammography System
|
|||
High voltage generator
|
Type
|
High frequency inverter type
|
|
X-ray tube voltage
|
20-40kV(step 0.5kV)
|
||
X-ray tube current
|
10mA-160mA
|
||
Maximum output power
|
5kW(40kV,125mA)
|
||
Nominal power
|
4.8kW(30kV,160mA,0.1s)
|
||
Radiograph time
|
0.005-10s
|
||
mAs
|
0.5mAS~560mAs
|
||
Minimum mAs
|
0.5mAs(50ms 10mA)
|
||
Maximun mAs
|
560mAs
|
||
Minimum mAs in AEC
|
6.3mAs
|
||
Exposure control
|
manual ,semi-auto,auto
|
||
Radiating mode
|
cross-ventilation
|
||
X-ray tube
|
|
XM15
|
HMT-113
|
|
small focus/big focus
|
small focus/big focus
|
|
Focus size
|
0.1mm/0.3mm
|
0.1mm/0.3mm
|
|
Maximum anode heat capacity
|
225kJ(300kHU)
|
225kJ(300kHU)
|
|
Maximum heat dissipation
|
500W
|
500W
|
|
Anode cooling mode
|
oil cooling
|
oil cooling
|
|
Nominal voltage
|
40kV
|
40kV
|
|
Target angle
|
15°
|
10°/16°
|
|
Target material
|
Mo
|
Mo
|
|
Radiation field
|
18cm-50cm
|
18cm-50cm
|
|
Inherent filtration
|
0.5mmBe
|
0.5mmBe
|
|
Additional filtration
|
0.03mmMo
|
/
|
|
Anode speed
|
low-speed3000r/min; high-speed 10000r/min
|
low-speed3000r/min; high-speed 10000r/min
|
|
Anode input power
|
1.15-2KW / 4.8-9kW
|
0.7-1.35kW / 2.9-4.9kW
|
|
X-ray tube socket (C339v)
|
|
|
|
Additional filtration
|
0.03mmMo
|
||
Weight
|
13kg
|
||
C Arm frame
|
|
|
|
vertical range of motion
|
630-1400mm,deviation±5%
|
||
Rotation range of motion
|
+195°~-155°Angular division value 1°,deviation±2%
|
||
FID
|
660mm,deviation±5%
|
||
3D tomographic axial rotation
|
+30°~-30°deviation±2%
|
||
Rotate Mode
|
power-drive double speed rotate
|
||
|
Preset angle and memory function
|
||
Oppression plate
|
range of motion
|
245mm,deviation±10%
|
|
range of motion in 1.5 times platfo
|
150mm,deviation±10%
|
||
range of motion in 1.8 times platfo
|
75mm,deviation±10%
|
||
Operation mode
|
Manual/Auto
|
||
Pressure range
|
0-300N(Manual)150-200N(Auto)
|
||
FPD
|
|
AXS-2430
|
HMD-2430
|
Pixel unit
|
85um
|
85um
|
|
Effective pixel size
|
239.36*304.64mm
|
239.36*304.64mm
|
|
Pixel matirx
|
2816*3584 pixels
|
2816*3584 pixels
|
|
Spatial resolution
|
≥7 lp/mm
|
≥7 lp/mm
|
|
Cooling mode
|
fan cooling
|
fan cooling
|
|
Grayscale
|
≥16bit
|
≥16bit
|
|
DQE
|
≥50%
|
≥50%
|
|
Grid
|
|
|
|
Grid proportion
|
5:1
|
||
Size
|
338*264mm
|
||
line pair
|
36lp/cm
|
||
focus
|
66cm
|
||
Collimator (HMC-915)
|
|
|
|
Power supply
|
24V DC,2A
|
||
Radiation field
|
Min 5*5cm/Max 24*30cm
|
||
Radiograph platform
|
Field 24*30cm SID 660mm,deviation±5%
|
||
1.5 times radiograph platform
|
Field 16*20cm SID 440mm,deviation±5%
|
||
1.8 times radiograph platform
|
Field 13.6*16.7cm SID 366.7mm,deviation±5%
|
Collimator (HMC-915)
|
Inherent filtration
|
0.02mmAl
|
Additional filtration
|
0.03mmMo /0.025Rh
|
|
Illumination in max SID
|
>160Lux
|
|
Timer
|
Button/Auto,30s timer
|
|
Indicator light
|
LED
|
|
maximum radiation leakage
|
<40mRh
|
|
Light field switch
|
Auto
|
|
Weight
|
3.6kg
|
|
Image Acquisition Workstat
|
Screen size/resolution
|
19",Resolution 1280*1024,Contrat 1000:1
|
Memory
|
≥8G
|
|
Hard disk
|
≥320G
|
|
OS
|
Win7
|
|
Image format
|
DICOM3.0(can connect with RIS)
|
|
Image storage capacity
|
≥20000 images
|
|
|
mamual input patients' date or download form HIS/RIS through DICOM Worklist
|
|
|
CD-R writer (storage image)
|
|
|
Choosing Breast implants Function
|
|
|
Integrated design with High voltage generator, exposure specification setting in workstation
|
1).Does your product has any certificate?
Almost all our medical equipment is CE, ISO approved.
2).What's your payment terms?
We can accept T/T,Western Union, Money Gram, Paypal, cash etc
3).What's your product lead time?
Normally all our medical equipment are in stock, so we can deliver goods within 3 workdays after receipt payment.
4).How to maintenance the product if it has problem?
We have professional engineers,they will solve your problem by communicating with skype or phone. If the machine is totally broken and can not work, send it back to us and we will send a new one to you.
5).What's your shipping terms?
Foermed offers courier,air,sea shipping methods.Client can choose anyone depends on detail situation.