Customization: | Available |
---|---|
Material: | Oxford Cloth |
Application: | Hospital |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Defibrillator
|
|
Output
|
Biphasic Truncated exponential
|
Energy Sequence(child)
|
50J, 50J, 50J
|
EnergySequence (Adult)
|
150J, 200J, 200J
|
Charge Time
|
6 sec. to 150J 8 sec. to 200 J(New battery)
|
Analysis Time
|
6 sec.
|
The maximum time from the initiation of rhythm analy-sis to readiness for discharge with a new battery
|
Less than 10seconds
|
The maximum time from the initiation of rhythm analy-sis to readiness for discharge after 6 shocks
|
Less than 10 seconds
|
The maximum time from initially switchingpower on toreadiness for discharge
|
Less than18 seconds
|
Visual Prompts
|
LED prompts;Automatic Self-test
|
Controls
|
Four buttons: On/Off, Shock, Adult/child, CPR
|
Output Energy Accuracy
|
±10% into any impedance from 25 to 175Q
|
The Maximum Voltage
|
1400±50V
|
Output disabled when the patient impedance is outsidelimits
|
20Ω to 200Ω
|
Battery
|
|
Part No.
|
CR123A-4x2
|
Non-rechargeable
|
2.8 Ah
|
Capacity
|
120 discharges at 200 Joules
|
Shelf Life(25° C+15C)
|
5 years standby (after installation)
|
Physical
|
|
Dimensions
|
280×209 x 80 mm
|
Weight
|
2.5 kg
|
Operating Temperature
|
0°ºC to 40ºC
|
Operating Humidity
|
Relative humidity between 30% and 95%(non-condensing)
|
Storage Temperature (without battery)
|
-20° C to 55ºC
|
Storage Humidity(without battery)
|
Up to 93%(non-condensing)
|
Packing
|
Single package size: 40*30*17.8cm 6 unit per carton, package size: 62.5*42.5*56. 5cm, total gross weight 21 KG
|