Customization: | Available |
---|---|
Customized: | Non-Customized |
Certification: | CE, FDA, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Classic style Double Pump Hemodislysis Machine
Model | YJ-D2000 |
Power | 1500W |
Working environment | temperature 10ºC ~30ºC (at relative humidity of no more than 70%.) |
Water input pressure | 0.15 MPa ~0.6 MPa |
Water input temperature | 10ºC~30ºC |
Machine Size | 380mm×400mm×1380mm (L*W*H) |
Package Size | 680*550*1680mm |
Net/Gross Weight | 88KG/93kg |
Dialysate | Dialysate temperature: preset range 34.0ºC~39.0ºC Dialysate flux: 300~800 ml/min Dialysate concentration: 12.1 mS/cm ~16.0 ms/cm, ±0.1 ms/cm Dialysate mixing ratio: can set variety ratio. UF rate Flow range: 0 ml/h ~4000 ml/h Resolution ratio: 1ml Precision: ±30 ml/h |
Extracorporeal Part | Venous pressure: -180 mmHg ~+600 mmHg, ±10 mmHg Arterial pressure : -380 mmHg ~+400 mmHg, ±10 mmHg TMP pressure : -180 mmHg ~+600 mmHg, ±20 mmHg Blood pump flow range: 20 ml/min ~400 ml/min(diameter:Ф6 mm) Spare pump flow range: 30 ml/min ~600 ml/min(diameter:Ф8 mm) Resolution ratio: 1 ml Precision: error range ±10ml or 10% of reading |
Heparin Pump | Syringe size: 20, 30, 50 ml Flow range: 0 ml/h ~10 ml/h Resolution ratio: 0.1ml Precision: ±5% |
Sanitize | 1.Hot decalcification : Time: about 20 minutes Temperature: 30~60ºC, 500ml/min. 2.Chemical disinfection : Time: about 45 minutes Temperature: 30~40ºC, 500ml/min. 3. Heat disinfection : Time: about 60 minutes Temperature: >85ºC, 300ml/min. Storage Environment Storage temperature should be between 5ºC~40ºC, at relative humidity of no more than 80%. |
Monitoring System | ialysate temperature: preset range 34.0ºC~39.0ºC, ±0.5ºC Blood leak detection: Photochromic Alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate Bubble detection: ultrasonic Alarm when a single air bubble volume is more than 200µl at 200ml/min blood flow Conductivity: acoustic-optic, ±0.5% |