Blood Gas Analyzer
- Accurate, reliable and maintenance-freeSelf-calibration for each test
Accurate results in approximately 5 minutes Dry chemistry method, no reagent pack required, no carry-over contamination - Lightweight and portable Rechargeable lithium battery: more than 50 tests Size: 235mm×210mm×160mm Weight: 3±0.5 kg (including battery) |
|
- Easy-to-use
Quick start tutorial 8-inch full HD touchscreen - Smart identification in cartridgeFeedback of cartridge insertion ldentification of the cartridge expiration date - Automatic quality control Regular QC reminders Power on Self-Test Dual quality control: electronic simulator and controls |
|
Blood Gas Test Cartridge
* 34 parameters of blood gases , electrolytes and metabolites
10 measured parameters: K+, Na+, Cl- , iCa2+, pH, pCO2 , pO2 , Glu, Lac, Hct
24 calculated parameters: cH+ (T), pH(T),pCO2 (T),pO2 (T),pO2 (A-a)(T),pO2 (a/A) (T),RI(T),pO2 (T)/FiO2 ,cH+,iCa2+(7.4),HCO3-act,HCO3-std,BE(B),BB(B),AG, sO2 (est),tHb(est),TCO2 ,pO2 (A-a),pO2 (a-A),RI,mOsm,pO2 /FiO2 |
Parameter
|
Accuracy
|
Range
|
Precision
|
K*
|
±3.0%
|
2.0-9.0 mmol/l
|
≤1.5%
|
Nat
|
±3.0%
|
100-180mmol/L
|
≤1.5%
|
CI
|
±3.0%
|
65-140 mmol/L
|
≤1.5%
|
iCa2*
|
≥1.00 mmo/L,±5.0%; <1.00 mmol/L, ±0.05 mmol/L
|
0.25-2.50 mmol/L
|
≤1.5%
|
pH
|
±0.04
|
6.500-8.000
|
≤0.02
|
pCO2
|
≥62.0mmHg,±8.0%; <62.0mmHg,±5.0 mmHg
|
10.0-150.0 mmHg
|
≥62.0mmHg, CV≤4.0%; <62.0mmHg, SD≤2.5 mmHg
|
pO2
|
≥50mmHg,±15.0%; <50mmHg,±7.5 mmHg
|
10-700 mmHg
|
≤5.0%
|
GIu
|
≥4.0 mmo/L, ±10.0%; <4.0 mmol/L, ±0.4 mmol/L
|
1.1-38.0 mmol/L
|
≥4.0 mmol/L, CV≤5.0%; <4.0 mmo/L, SD≤0.2 mmol/L
|
Lac
|
≥5.00 mmol/L,±12.0%; <5.00 mmo/L,±0.6 mmol/L
|
0.50-20.00 mmol/L
|
≥5.00 mmol/L, CV≤6.0%; <5.00 mmol/L, SD≤0.3 mmol/L
|
Hct
|
≥50%PCV, ±6.0%; <50%PCV, ±3%PCV
|
10%-70%PCV
|
≥50%PCV, CV≤3%; <50%PCV, SD≤1.5%PCV
|