Item
|
Specification
|
Magnet System
|
|
Magnet Type
|
Superconductive
|
Field Strength
|
1.5T
|
Shielding Method
|
Active
|
Shimming Method and Type
|
Active + Passive+ Dynamic
|
Magnet stability
|
s0.1ppm/h
|
Homogeneity (DSV, VRMS)
|
|
45cm
|
50.832ppm
|
40cm
|
50.323ppm
|
30cm
|
50.107ppm
|
20cm
|
s0.041ppm
|
10cm
|
s0.01ppm
|
Length of magnet(exclude cover)
|
157cm
|
Inner diameter of magnet
|
600mm
|
5 Gauss fringe fieldX,Y, Z axis)
|
≤2.5mx2.5mx4.0m
|
Total capacity of liquid helium (100% liquid helium full filled)
|
800L
|
Filling liquid helium period
|
24 Years
|
Liquid helium "zero" consumption technology
|
Yes
|
Gradlent System
|
|
Maximum gra dient field (single axis, invalid )
|
41mT/m
|
Maximum gradi ent slew rate( single axis, invalid)
|
187mT/m/ms
|
Minimum gradient rise time
|
0,22ms
|
Maximum gradient field and slew rate reached at the same time
|
Yes
|
Gradient cooling system
|
Water cooling
|
Full digital real-time transmit and receiving gradient control system
|