Customization: | Available |
---|---|
Certification: | CE, ISO13485 |
Brand Name: | Forermed |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Portable ICU Ventilator YJ-S1100
Application The ventilator makes a good performance in operation room,ICU department and emergency treatment.It used to assist or replace the spontaneous breathing for adult pediatric and neonatal more than 2kg. 25 years experience in Market-oriented ventilator make us professional and reliable, satisfying all your needs in ventilation. Due to the flexible configuration, good quality and competitive price, S1100A has soon become the superstar of market. Features 15" TFT touch screen displays the ventilation parameters, alarm information, and oscillograms, make every operation more easily. Multiple ventilation modes,suitable for ICU,emergency department and operation room etc: ( IPPV, A/C, PCV, SIMV,PSV, SPONT/CPAP, SIGH, MANUAL) 5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time. ETCO2 is standard configuration for this machine. Humidifier can heat and wet breathing gas, makes it comfortable for patient to breathe. Rapid oxygen supply,automatically offer high flow rate oxygen within two minutes Nebulization make medicine into small liquid and particle,easier and quickier for patient to breathe in. High temperature resistance breathing circuit is reusable and anti-pollution. |
Ventilation mode | IPPV, A/C, PCV, SIMV, PSV, APRV, PRVC, SPONT/CPAP, SIGH,MANUAL |
Ventilator parameter | |
Tidal volume (Vt) | 0 ~ 2000 mL |
Frequency (Freq) | 1 min ~ 100 min |
Oxygen concentration | 21 % ~ 100 % |
I:E | 4 : 1 ~ 1 : 8 |
PEEP | 0cmH2O ~ 40 cmH2O |
Pressure triggering sensitivity (Ptr) | -20 cmH2O ~ 20 cmH2O (Based on PEEP) |
Flow trigger sensitivity (Ftr) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 80 cmH2O |
Pressure support (PS) | 0 cmH2O ~ 80 cmH2O |
SIGH | 0 (off) 1/100 ~ 5/100 |
Apnea ventilation | OFF, 5 s ~ 60 s |
Pressure limit | 20 cmH2O ~ 100 cmH2O |
Monitoring parameter | |
Frequency(Freq) |
0 / min ~ 100 / min |
Tidal volume(Vt) |
0 mL ~ 2500 mL |
MV |
0 L/min ~ 99 L/min |
Airway pressure |
0 cmH2O ~ 100 cmH2O |
Dynamic lung compliance testing |
1 mL/cmH2O ~ 1000 mL /cmH2O |
ETCO2 concentration |
0 mm Hg ~ 152 mmHg (0 % ~ 20 %) |
Oxygen concentration |
15 % ~ 100 % |
Alarm and protection | |
AC power failure alarm |
Power failure or no connection |
Internal backup battery low voltage alarm |
≤ 11.3 ± 0.3 V |
No tidal volume |
No tidal volume within 6 s |
High minute volume alarm |
5 L/min ~ 99 L/min |
Low minute volume alarm |
1 L/min ~ 30 L/min |
High airway pressure alarm |
20 cmH2O ~ 100 cmH2O |
Low airway pressure alarm |
0 cmH2O ~ 20 cmH2O |
High oxygen concentration alarm |
19 % ~ 100 % |
Low oxygen concentration alarm |
18 % ~ 99 % |
Continuous pressure alarm |
(PEEP + 1.5 cmH2O) over 16 s |
Suffocation warning |
5 ~ 60 s |
Fan error |
Show on screen |
Oxygen deficit |
Show on screen |
The maximum limited pressure |
< 12.5 kPa |
Working condition | |
Gas source | O2, Air |
Pressure | 280 kPa - 600 kPa |
Voltage | 220 V ± 22 V |
Power frequency | 50 Hz ± 1 Hz |
Input power | 900 VA (With air compressor) 250 VA (Without air compressor) |
Oscillogram | P-T(Pressure-Time) F-T(Flow-Time) V-T(Volume-Time) ETCO2 -T(End-Tidal CO2-Time) P-V Loop(Pressure-Volume Loop) F-V Loop(Flow-Volume Loop) P-F Loop(Pressure-Flow Loop) |
Packing size | Main components: L 560 * W 560 * H 605 mm G.W. : 40 KG, N.W. : 17 KG Air compressor: L 670* W 700 * H 1160 mm G.W. : 84 KG, N.W. : 46.2 KG |