Dialysis machine YJ-5000
Features:
The blood purification equipment with full functions: CVVHDF, CPFA, DFPP, MARS, FPSA, HP … Full Range Of Blood Purification Equipments Citrate anti-coagulation and double input channel for substitution fluid flow |
Clinical advantages of citrate anticoagulation:
* All treatment modes can adopt the Citrate anti-coagulation.
* Seperate citrate infusion pump and weighing scale.
* Control the citrate flow accurately.
Automatic injection by proportion,achieve the safe topical citrate
anticoagulation. |
Product parameters
|
|
Basic Information
|
|
Dimension
|
Height 1720 mm/Width 635 mm/Thickness 578 mm
|
Weight
|
About 75Kg
|
Electrical Parameters
|
|
Voltage
|
-110/220V
|
Frequency
|
50/60Hz
|
Power
|
500VA
|
Backup Battery Capacity
|
2×12V/4Ah
|
Pressure Monitoring
|
|
Venous Pressure Monitoring(Vp)
|
-500mmHg~+700mmHg Accuracy:±5mmHg |
Arterial Pressure Monitoring(Ap)
|
-500mmHg~+700mmHg Accuracy:±5mmHg |
Transmembrane Pressure Monitoring(Tmp)
|
-500mmHg~+700mmHg Accuracy:±5mmHg |
Primary Membrane External Pressure Monitoring(Pm1)
|
Accuracy:±5mmHg -500mmHg~+700mmHg |
Secondary Membrane Pressure Monitoring(Pm2)
|
-500mmHg~+700mmHg Accuracy:±5mmHg |
Liquid Balance Control
|
|
ScaleNumber
|
4
|
Rehydration Scale
|
Accuracy;±5gor ±0.1% Okg~30kg |
Waste Liquid Scale
|
Okg~30kg Accuracy:±5gor ±0.1% |
Auxiliary Rehydration Scalel
|
Okg~12kg Accuracy;±5gor±0.1% |
Flow Monitoring
|
|
Blood pump flow(BP)
|
0mL/min,30mL/min~600mL/min Accuracy: set value±10% |
LP1
|
0mL/min,1ml/min~250ml/min Accuracy:±0.1mL/min or±5% ofsetvalue |
Waste Liquid Pump Flow(UFP) |
0mL/min,1mL/min~250mL/min Accuracy:±0.1mL/min or±5% ofsetvalue |
LP2
|
0mL/min,1mL/min ~250mL/min Accuracy:±0.1mL/min or±5% ofsetvalue |
Function Pump Flow(FP) |
0mL/min,1mL/min~250mL/min Accuracy:±0.1mL/minor±5% ofsetvalue |
Infusion pump flow
|
|
Flow Adjustable Range
|
0mL/h~10mL/h
|
Error Range
|
0.2mL/hor±5%
|
Rapid Injection Flow
|
10mL/h~2000mL/h
|
Syringe Specification
|
10mL/20mL/30mL/50ml
|
Dehydration Error
|
<±20mL/h
|
Liquid Balance Control
|
100ml
|
Substitution fluid flow temperature control
|
|
Control Range
|
33ºC~ 40ºC
|
Accuracy
|
Control error ±1° C
|
Safety Monitoring
|
|
Blood Leakage Monitoring |
≤ 0.35mL/min(HCT 32%)
|
Air Monitoring
|
>0.02mL bubbles
|
Note: The above parameters are YJ-5000 series, Specific parameters are subject to the operating manual.
|