Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station.
Performance Specifications
|
|
Dimension and Weight
|
|
Dimension
|
360mm*310mm*120mm
|
Weight
|
5.3kg(excluding accessories)
|
Power Supply
|
Voltage AC100~240V, 50/60HZ
|
Display
|
12.1" color TFT LED (0-90° rotation) Resolution 1024*768 pixels
|
Battery (Built-in)
|
Type Rechargeable lithium battery 14.8V/2200mAh
|
Charge Cycle
|
≥500 times
|
Working time
|
2 hours
|
Recorder (Built-in)
|
|
Method
|
Thermal printer
|
Paper
|
Z-fold, thermosensitive
|
Paper width
|
112 mm (4.41 in)/150 mm (5.91 in)
|
Printing speed
|
1/2/3cm/min
|
Trace
|
Max. 4 tracks
|
-Recording way
|
Real-time Recording, Review Printing
|
Alarm
|
|
Level
|
Low, medium and high
|
Indication
|
Auditory, Visual, Delay of FHR Alarm
|
Input Device
|
|
-Touch screen (option)
|
|
-Knob (Standard)
|
|
-Keypad input (Standard)
|
|
Mouse/Keyboard input (option)
|
|
-USB port for patient data/NST report backup (option)
|
|
System Output & Extensible Interface
|
|
-Ethernet Network
|
1 standard RJ45 socket
|
-USB Port
|
2
|
-Video Output
|
1 HDMI port (option)
|
Trend & Reviewing
|
|
-Cardiotocography (CTG)
|
|
Waveform Review
|
8640 hours
|
-Vital Signs
|
|
Trend
|
2160 hours
|
Waveform Review
|
2 hours
|
NIBP Review
|
2000 groups
|
-Alarm Review
|
2000 groups
|
Safety
|
|
-IEC60601-1 Approved, CE marking according to MDD93/42/EEC
|
|
-With reference to RoHS Directive 2011/65/EU recasting
|
|
-Ultrasound (FHR)
|
|
-Transducer type
|
12-Crystal ultrasound transducer
|
-Measure method
|
Wide beam pulse doppler
|
-FHR Measurement range
|
|
30 to 240 bpm (US Standard)
|
|
50 to 210 bpm (Europe Standard)
|
|
Accuracy
|
±1 bpm
|
Intensity
|
<3 mW/cm2
|
-Ultrasound frequency
|
1MHz±3%
|
Resolution
|
1 bpm
|
TOCO
|
|
-Measurement method
|
Strain gauge sensor element
|
-Measure range
|
0 ~ 100 units
|
Resolution
|
1 unit
|
-Nonlinear deviation
|
≤±8%
|
-Baseline setting
|
0,5,10,15, 20 can be selected
|
-Auto offset correction
|
3 second after connecting the transducer
|
-Auto zero adjust
|
TOCO value is set to zero negative following a negative Measure value for 5 seconds
|
Fetal Movement Recording
|
|
-Automatic fetal movement (AFM)
|
pulsed doppler ultrasound
|
-Automatic fetal movement autograph
|
|
-Manual fetal movement marker (MFM)
|
|
Maternal ECG
|
|
-Lead mode
|
3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
|
A63Gain
|
2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, Auto
|
Sweep speed
|
6.25mm/s, 12.5mm/s, 25mm/s, 50 mm/s
|
Accuracy
|
±1bpm/ ±1%, whichever is greater
|
Resolution
|
1 bpm
|
Leakage Current < 10 μA
|
|
CMRR
|
Diagnostic Mode >89 dB
|
|
Monitor Mode >105 dB
|
|
Surgery Mode >105 dB
|
Bandwidth
|
Surgery 1 ~ 25 Hz
|
|
Monitor 0.5 ~ 40 Hz
|
|
Diagnostic 0.05 ~ 120 Hz
|
Maternal Heart Rate
|
|
-Measure range
|
Adult:0bpm; 10 ~ 300 bpm
|
|
Neonate:0bpm; 10 ~ 350 bpm
|
Resolution
|
1 bpm
|
Accuracy
|
± 1% or ± 1bpm, whichever is greater
|
ST Measurement
|
|
Range
|
-2.0 ~ +2.0 mV
|
Accuracy
|
-0.8mV~+0.8mV:±0.02mV or ±10%, whichever is greater
|
-Other range
|
unspecified
|
Resolution
|
0.01mV
|
Maternal Respiration
|
|
Method
|
Impedance between RA-LL, RA-LA
|
Gain
|
×0.25, ×0.50, ×1, ×2, ×4
|
-Respiration Rate
|
0bpm, 6 ~ 150 bpm
|
Resolution
|
1 bpm
|
Accuracy
|
±2 bpm or ±2%, whichever is greater
|
-Apnea Alarm
|
10 ~ 60 s
|
Maternal SpO2 (Digital Technic)
|
|
-Measurement Range
|
0 ~ 100 %
|
Resolution
|
1%
|
Accuracy
|
±2% (70% ~ 100%)
|
±3 % (35% ~ 69%)
|
|
Unspecified (0 ~ 34%)
|
|
Pulse Rate
|
|
Range
|
25~250bpm
|
Accuracy
|
±1% or ±1 bpm, whichever is greater
|
Resolution
|
1bpm
|
Maternal NIBP (MNIBP)
|
|
Method
|
Oscillometric
|
-Measure mode Manual, Auto, STAT
|
|
-Measure Interval in AUTO Mode 1~480 min
|
|
-STAT mode cycle time
|
|
-Keep 5 minutes, at 5 seconds interval
|
|
-Measure and Alarm Range
|
|
Systolic
|
30~270 mmHg
|
Diastolic
|
10~220 mmHg
|
Mean
|
20~235 mmHg
|
-Static pressure accuracy ±3mmHg
|
|
Resolution
|
1mmHg
|
Accuracy
|
Maximum Mean error ±5mmHg
|
Maximum Standard deviation 8mmHg
|
|
-Over pressure Protection
|
Dual protection via software & hardware
|
Maternal Temperature
|
|
Technique
|
Thermistor probe (2.25K)
|
-Measuring and Alarm Range
|
0.0 ºC~ 50 ºC (32ºF~ 122ºF)
|
Unit
|
Celsius (ºC), Fahrenheit ( ºF)
|
Resolution
|
0.1ºC
|
Accuracy
|
±0.1ºC
|
NST Report/FHR Interpretation (Option)
|
|
-Based on Fischer Evaluation Method (monitoring time≥20 Min)
|
|
-Average Baseline FHR /Number of Acceleration /Number of Deceleration /Number of Small Acceleration /Amplitude Variation /Number
of PV |
|
-Automatic Score Result
|
Score 8-10 (Excellent)
|
|
Score5-7 (Suspicious)
|
|
Score ≤4 ( Fetus Hypoxia)
|