Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | Digital Mammography Machine X-ray for Breast |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specification |
|
No. |
Parameters and functions |
1 |
System Overview (Basic Requirement) |
1.1 |
Clinical application: dedicated to human breast tissue X-ray photography to obtain tissue images for breast disease diagnosis and breast cancer screening; |
2 |
Frame Structure |
2.1 |
Rack structure: The C arm is developed and produced by manufacturer. (Possess all independent intellectual property rights) |
2.2 |
C arm rotation range: -155°~ +195° |
2.3 |
C arm vertical movement range: 630 ~ 1440mm; |
2.4 |
C arm can be locked anywhere in the full stroke |
2.5 |
The display on the rack can display the current position and the position information to be photographed, and has a one-key switch function to the preset position; |
2.6 |
FID: 660mm; |
3 |
High Voltage Generator Component |
3.1 |
Maximum output power: 5kW |
3.2 |
Output voltage range: 20 ~ 40 kV |
3.3 |
Output current range: 10 ~ 160 mA |
3.4 |
Loading time range: 0.005s ~ 10s |
3.5 |
Current time product range: 0.5 ~ 560mAs; |
3.6 |
With exposure control system, support for one-key exposure or hand brake exposure |
4 |
X-ray Tube Component |
4.1 |
Double speed rotating anode |
4.2 |
Nominal tube voltage: 40 KV; |
4.3 |
Anode heat capacity: 300 kHU; |
4.4 |
Anode rotating speed: 10000 rpm |
4.4 |
Anode material: molybdenum |
4.5 |
Nominal focus size: small focus 0.1mm, large focus 0.3mm; |
5 |
Flat Panel Detector Component |
5.1 |
Imaging material: amorphous silicon |
5.2 |
Imaging size: 240 × 300mm; |
5.3 |
Conventional spatial resolution: 7.01 p/mm; |
5.4 |
Pixel matrix: 2816 × 3584mm pixels; |
5.5 |
A/D output: 16 bits; |
5.6 |
Cooling method: fan cooling |
5.7 |
Weight: 3.5 kg |
5.8 |
Pixel size: 85 μm |