Customization: | Available |
---|---|
Type: | Syringe Pump |
Kind: | Syringe Pump |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Injection Rate Range
|
50ml syringe : 0.1~999.9ml/h (step :0.1ml/h)
1000~1800ml/h (step : 1ml/h)
30ml syringe :0.1~900ml/h (step :0.1ml/h) 20ml syringe :0.1~600ml/h (step :0.1ml/h) 10ml syringe :0.1~300ml/h (step :0.1ml/h) 5ml syringe :0.1~150ml/h (step :0.1ml/h) |
Purge Rate
|
50ml syringe: 1800ml/h (Bolus rate: 1200ml/h) 30ml syringe:900ml/h (Bolus rate: 600ml/h) 20ml syringe:600ml/h (Bolus rate: 400ml/h) 10ml syringe:300ml/h (Bolus rate: 200ml/h) 5 ml syringe:150ml/h (Bolus rate: 100ml/h) |
Mechanical Precision
|
Within±2%
|
Preset Volume Range
|
0.1~1999.9ml
|
Occlusion Alarm Threshold
|
High: 800mmHg±200mmHg (106.7kPa± 26.7kPa) Medium: 500mmHg ± 100mmHg (66.7kPa± 13.3kPa) Low: 300mmHg ± 100mmHg (40.7kPa± 13.3kPa) |
KVO Rate
|
0.1~5ml/h (step: 0.1 ml/h)
|
Fuse
|
F1AL/250V 2 pcs (installed inside pump) |
Power Supply
|
100~220V, 50Hz, Built in Rechargeable Li-ion battery, 11.1V, Capacity ≥2000mAh; by charging 10 hours, it can work at 5ml/h (GB 9706.27-2005 medium rate) for 4 hours |
Environment Requirement
|
Operation:
a) Ambient temperature: +5ºC~+40ºC
b) Relative humidity: 20%~ 90% c) Atmosphere Pressure: 86.0~106.0 (kPa) |
Transportation & Storage:
a) Ambient temperature: -30ºC~+55ºC
b) Relative humidity: ≤95% c) Atmosphere Pressure: 50.1~106.0 (kPa) |
|
Alarm
|
Near, Injection finished, Preset dose injection finished, Occlusion, Syringe off, Syringe dislocated, Low battery, AC power off
|