Customization: | Available |
---|---|
Certification: | CE |
Product Name: | Magnetic Resonance Imaging |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
CT Scan Unit Digital X-ray Machine YJ-CT32
*An imventive patent an Precise Tomography Technique
* The original 32-slice spectral imaging technique * The independent R&D on Campo past-processing workstation(CPW)
The original 32-slice spectral imaging technique
Far the first time, we have applied the spectral imaging tochnology to the 32-ayer CT, currently, this techn ology is only applied for high-end CT.To diagnose lesions and tissues in aroas where other CTscanners are difficult to achieve, improving the diagnostic accuracy and safety, and the success rate of imaging diagnosis is greatly improved.
The independent R&D on Campo post-processing workstation(CPW) Fully ment the medical image connectivity standard, have powerful 20 reading tablets, free film printing templates, professional patient report templates; allow multi-level reconstruction of MPR, as well as cmossesection, sagittal section, coronal section, ablique cut section and curve reconstruction; Perfect for 3d visualization, high quality, stereascopic dispbay (VR), maximum intensity projection (MIP), minimum intensity projection (MiniP), average intensity projection(AIP) and sagittal section display(SSQ,etc. An inventive patent on Precise Tomography Technique Precision Raxial tomagraphy patent techindlog, the world's only longitudina! sampling, innovative way of scamning incmis pasents conbined with a cousite racoustruction algonithn, tron the daa sa mpling and imaga rnaconstruction af two fidda sogether for longitudieal uftra high-resollution in apa racomstruction, significantly incrmase the detection sate of smaillesions |
S/N
|
Item
|
Specifications
|
1
|
Gantry
|
|
1.1
|
Slip Ring
|
Low-voltage slip ring
|
*1 2
|
Aperture Size
|
715mm
|
1.3
|
Tiltangulation Range
|
±30°
|
1.4 |
Laser Positioning System |
Provided, provide sagittal, coronal and internal and external cross-sectional positioning lines
|
1.5
|
Scanning Collimation
|
2x1. 1mm;8x0.55mm;16x0.55mm;16x1. 1m m
|
1.6
|
Sampling Rate
|
1800 view/ cycle
|
1.7
|
Min Scan Time (s/360°)
|
0.72s
|
1 8
|
Max FOV (Field of View)
|
500mm
|
1.9
|
Data Transmission Mode
|
Digital transmission
|
1. 10
|
Cooling-down Method
|
Air cooling heat dissipation
|
2
|
Detector
|
|
2.1
|
Detector Type
|
GOS ceramic solid detector
|
2.2
|
Detector Number
|
888x24
|
2 3
|
Detector Row
|
24
|
2.4
|
Detector Channel Per Row
|
888
|
*2.5
|
Detector Thickness Per Row
|
0.55mm
|
|
|
|
3
|
X-ray System
|
|
*3. 1
|
Focal Point
|
Bifocus
|
3.2
|
Focal Spot Size
|
0.7mmx0.8mm; 1.2mmx1.4mm
|
*3.3
|
Heat Storage Capacity
|
3.5 MHu
|
3.4
|
Max Anode Cooling Rate
|
8750W (12250 HU/sec)
|
3.5
|
Tube Life
|
100000 second exposure
|
3.6
|
Max Output of Tube Voltage
|
140kV
|
*3.7
|
Min Output of Tube Voltage
|
60kV
|
3.8
|
Output Voltage
|
60kV, 70kV, 80kV, 100kV, 120kV, 140kV
|
3.9
|
Max Output of Tube Current
|
350mA
|
*3. 10
|
Min Output of Tube Current
|
10mA
|
*3. 11
|
Max Power
|
42kW
|
3. 12
|
Max Continuous Exposure Time
|
100s
|
4
|
Scanner Table
|
|
4.1
|
Horizontal Movement Range
|
0 ~ 1725mm
|
4.2
|
Max Scan Range
|
1650mm
|
*4.3
|
Max Horizontal Speed Range
|
300mm/s
|
4.4
|
Repetitive Positioning Accuracy
|
±0.25mm
|
4.5
|
Min Vertical Range
|
450mm
|
4.6
|
Vertical Lift Range
|
470mm
|
*4.7
|
Max Load Capacity
|
205kg
|
5
|
Scanning and image reconstruction system
|
|
5.1 |
Scan Mode
|
Tomoscan,Helical Scan, Precision Tomoscan, Dual Energy Scan (Optional ) |
*5.2
|
Min Image Thickness
|
0.275mm
|
*5.3
|
Recon Matrix
|
512×512,768×768, 1024×1024
|
*5.4
|
Max Recon Matrix
|
1024×1024
|
5.5
|
Max Continuous Scan Length
|
1650mm
|
5.6
|
Max Scan Field of View
|
500mm
|
5.7
|
Max Recon Field of View
|
500mm
|
5.8
|
Min Recon Field of View
|
50mm
|
5.9
|
Spatial Resolution
|
16.6LP/cm@MTF10% 21LP/cm@MTF0% |
5. 10
|
Low Density Resolution
|
|
5 . 1 1
|
CT Value Display Range
|
±16383
|
5. 12
|
Display Gray Scale
|
256, 24bit color
|
5. 13 |
Thin-layer Scanning of Different Thick Layers Reconstruction Function
|
Provided |
5. 14
|
Iterative Reconstruction
|
Provided, P-IR dual-domain iterative de-noising reconstruction technology, real
|