Customization: | Available |
---|---|
Type: | Hematology Analyzer |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
• Microprocessor based servo controlled temperature system
• Three modes: pre-warm, manual or automatic
• Set temperature and skin temperature failure,over temperature.
• APGAR timer to record therapy time
• Convenient X-ray cassette tray under infant mattress
• Heater head can be adjusted ±90° horizontally
• Inclination of infant bed is ajustable
• LED observe light angle is adjustable
• Transparent protector can be folded down, with two infusion seals
• RS232 connector
• Optional parts; bigger infant bed, low pressure aspirator, LED phototherapy unit,
internal/external infant resuscitor (T pieces), air and oxygen blender, gel mattress, oxygen supply system,
electric height adjustable, air temperature & count down timer, infant head fixing unit, infant oxygen hood
Temperature control range
|
25°C -37°C
|
Temperature override mode to
|
38°C
|
Temperature sensor accuracy
|
≤0.3°C
|
Mattress surface temperature uniformity
|
≤2.0°C
|
Temperature rise time
|
≤45mins
|
Heat output indicate
|
0-100% in 10%
|
Radiant box horizon angle adjust range
|
0-90°adjustable in step-less
|
Bed tilt adjustment
|
continues adjustable in ±10°
|
Mattress tray size
|
1700px*1350px
|
Mattress tray Max. load
|
10kg
|
Transfusion pole Max. load
|
2kg
|
Shelf Max. load
|
2kg
|
YJ-BIW100 Infant Radiant Warmer
Brief introduction:
1. Microprocessor based servo controlled temperature system.
2.The microprocessor-controlled unit allows for three settings: pre-warm, manual or automatic.
3.Alarm functions for safety includes:
Power failure;
Temperature deviation;
Temperature sensor failure;
Over temperature.
1. APGAR timer to record therapy time
2. Convenient X-ray cassette under infant bed
3. Heater head can be adjusted +90º horizontally
4. Tilt angle of infant bed is adjustable
5. Observe light angle is adjustable
6. Transparent protector folded
7. Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source,the heating is quick and life-span is long
8. Alarms are audible and visual
9. A second thermal cut-out function for more safety
10. Skin temperature sensor failure protect function avoid over temperature
11. Temperature deviation can be amended in front panel directly
Standard configuration:
Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette , Observe light , I.V.pole, skin temperature sensor, tray, mattress, transparent protector, castors.
Optional parts:
Drawer;
RS-232(485) connector;
Technical data:
1.Power supply : 110/220V 60/50Hz
2.Power input: ≤1200VA
3.Operating condition:
An ambient Temperature:18°C -30°C
An ambient Relative Humidity: 30%--75%RH
Atmospheric Pressure: 700~1060 hPa
An ambient Air Velocity < 0.3 m/s
4. Temperature control: 32°C~38°C
5.Temperature control accuracy: ≤0.5°C
6.Mattress temperature uniformity : < 2°C
7.Skin temperature sensor accuracy: + 0.3°C
8.Warm-up time: ≤30min
9. Infant bed tilt angle:+ 10°
10.APGAR timer range: ≤24h
11.Heater head horizontal: ±90°
12.Mattress size: 67cm × 54cm
13.Distance from heater to mattress:80cm
Transport and storage:
Environment Temperature:-40°C ~ +55°C
Environment Relative Humidity: ≤95%
Atmospheric Pressure: 500~ 1060hPa
Package:
Each unit is packed one case
Size:120cm× 70cm × 96cm
Gross weight: ≤90kg