Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Kind: | Orthopedics |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Main Body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of Application: | Adult, Child |
Gas Source | O2: 0.28~0.6MPa, N2O: 0.28~0.6MPa |
Flow Meter | O2: 0.05~1.0L/min,1.1~10L/min N2O: 0.05~1.0L/min,1.~10L/min |
O2-N2O Linkage and N2O Stopper | When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.16±0.02Kpa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply | 25~75L/min |
Low Oxygen Pressure Alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. |
Respiratory Bellows | Bellows for adults and children Tidal volume range:0~1500 ml |
Respiratory Circuit | Working mode: All-close, Semi-close, Semi-open APL: ≤12.5 kPa |
Light | Equipped with LED illumination, easy for observation |
ACGO Port | With ACGO port to connect with T circuit or BAIN circuit for open-operation. |
AGSS Port | AGSS exhaust gas collection system can be installed. |
Display Mode | High-definition 10.1”TFT color LCD touch screen display, easy and convenient operation |
Ventilation Mode | IPPV, SIPPV, SIMV, PEEP, SIGH, MANUAL |
Ventilation Function | PEEP,SIGH, IRV |
Ventilation Parameters | Tidal volume:50~1500 ml Rate:2~99 bpm SIMV rate:2~20 bpm I:E:2:1~1:8 Inspiratory trigger pressure: -10~20 cmH2O PEEP:2~20 cmH2O Pressure range:5~60 cmH2O SIGH:1.5 times the inspiratory time/ 100 times breath |
Parameters For Ventilation Monitoring | Tidal Volume, Ventilation Volume, IPPV Rate, SIMV Rate, I:E, Peak Pressure of Airway, Pressure-Time Waveform, Flow Rate-Time Waveform, PEEP, Inspiratory Trigger Pressure (Average Pressure, FiO2 Oxygen Concentration, Battery Level) |
Security Alarm System | Oxygen Concentration Alarm: Upper limit setting range:21%~99%,Low limit setting range:18%~80% Airway Pressure Alarm: Upper limit setting range:5~60 cmH2O,Low limit setting range:0~50cmH2O Per-minute Ventilation Volume Alarm: Upper limit setting range:1~20L/min,Low limit setting range: 0~19L/min Tidal Volume alarm: Upper limit setting range:100~1500ml,Low limit setting range 0~1400ml |