6 m³ EO Sterilizer
serial number | name | Description of technical parameters |
1. | Model and power | H6.0 |
12KVA | ||
2. | Inner container size | 3000*1200*1700mm (allowable deviation ± 10%) |
3. | outline dimension | 4080*1490*2240mm (allowable deviation ± 10% |
4. | weight | About3.0 tons (net weight) |
5. | working pressure | -70kpa ~ 60kpa for 60min, and the leakage rate is less than 0.01kpa/min. |
6. | The sterilization temperature |
35ºC-55ºC |
7. | Sterilization humidity |
30%-90% |
8. | Door opening mode | Fan-shaped or movable doors are optional |
9. | Inner material | Adopt 304 stainless steel national standard steel plate with thickness ≥6mm,and the whole plate is welded by argon arc welding. Rugged and durable, the service life of the inner container exceeds 10 years. |
10. | External material | Using high quality 304 stainless steel wire drawing plate = 1.2mm. |
11. | mode of heating | High-quality widened PTC silica gel plate is used for heating on six sides, which has independent temperature control in all directions, good thermal conductivity and uniform temperature, and the inner container is provided with double protection of sterilization |
12. | Sterilizer door | Adopt single door structure, pneumatic single door seal (chute automatically opens and closes the door smoothly). In case of failure in use, it should always be locked and sealed. |
Sealing mode: silicone rubber external inflatable seal. | ||
13. | Monitor instrument | The pressure, humidity and temperature monitoring device can automatically monitor in real time, and the accuracy and error are 1%. |
14. | Humidifying system | The steam pipeline is made of SUS304. Equipped with automatic steam generator and corresponding pipelines to form an independent steam injection pipeline and filter (sterilization level). The steam pipeline is made of SUS304. |
Automatic real-time monitoring, alarm when the humidity exceeds the set value, and humidity error in the cabinet: 5% RH. |