Customization: | Available |
---|---|
Product Name: | Digital Electrocardiograph |
Model Name: | Yj-ECG3 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Lead
|
Standard 12 leads
|
Acquisition Mode
|
Simultaneous 12 leads acquisition
|
Measurement Range
|
±5mVpp
|
Input circuit
|
Floating; Protection circuit against Defibrillator effect
|
Input Impedance
|
≥50MΩ
|
Input circuit current
|
<0.05µA
|
Record mode
|
Automatic: 3CH×4+1R, 3CH×4, 3CHx2+2CHx3,3CHx2+2CHx3 +1R,6CHX2 Manual: 3CH, 2CH, 3CH+1R, 2CH+1R; Rhythm: Any lead selectable |
Filter
|
EMG Filter: 25 Hz / 30 Hz / 40Hz/75 Hz / 100 Hz / 150Hz
DFT Filter: 0.05 Hz/ 0.15 Hz
AC Filter: 50 Hz / 60Hz
|
CMRR
|
>100dB
|
Patient leak current
|
<10µA (220V-240V)
|
Frequency Response
|
0.05Hz~150Hz (-3dB)
|
Sensitivity
|
2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV (error:±5%)
|
Anti-baseline Drift
|
Automatic
|
Time constant
|
≥3.2s
|
Noise level |
<15µVp-p
|
Paper speed:
|
12.5 mm/s, 25 mm/s, 50 mm/s
|
Recording mode
|
Thermal printing system
Paper specification:Roll 80mmx20M
Safety Standard:IEC I/CF
|
Sample Rate
|
Normal: 1000sps/channel
|
Power Supply
|
AC: 100~240V, 50/60Hz, 30VA~100VA
DC: 14.8V/2200mAh, built-in lithium battery
|