YJ-DXP03 Portable Dental X Ray Machine
Voltage and current:
60KV-2mA(+5%)
More suitable oral film parameters, less output radiation, without large dosescan also take clear photos
Compared with common competitive products in the market, the volume isreduced by 25%, the design is compact, the portability is enhanced, and theoperation is more
convenient.
Built by a famous design team, the product is ergonomic, the overall edge andarc treatment, more comfortable to hold, long-term shooting will not be tired.
High-density battery, strong endurance, no false power, shooting worry-free.
Inverter frequency:
The inverter frequency is 70kHz, and the high-frequency inverter technologymakes the dose more stable, less invalid rays, and the image clearer.
The solid state glue filing process in the ball bag has no risk of oil leakage and is more efective in isolating radiation;There is no chemical harm after falling, and the service life is longer.
Security:
360-degree lead wrap ensures no dead space for radiation leakage.
Display screen:
LCD color screen, clear interface, functional Ul interaction more colorful, more comprehensive functions, easy to get started.
Technical specification
Serial number | Name | Parameter value | |
1 |
Power Adapter |
Input Voltage | Uniphase AC 100~240V,±10% |
Supply Frequency | 50/60Hz,±1Hz | ||
Input Power | 286W | ||
2 | Output Power | 25.2W | |
3 | Charge voltage | DC 25.2V | |
4 | Charge current | 1.0A | |
5 | Nominal value of the focal point | 0.8mm | |
6 | Anode angle | 19º | |
7 | Available maximum tube voltage at 2mA | 60kV | |
8 | Maximum output electric power | 0.12kW | |
9 | Maximum loading factor combination | 60kV ,2mA ,2s | |
10 | Repeatability of radiation output | No more than 0.5 | |
11 | Tube voltage (The deviation is within the range of ±10%) | 60kV | |
12 | Tube current (The deviation is within the range of ±20%) | 2mA | |
13 | Running mode | Discontinuous operation, maximum excitation (open) time : minimum excitation (close) time = 1:30 |
|
14 | Total filtration of X-ray tube assembly | >2.5mm Al | |
15 | Intrinsic filtration of X-ray tube assembly | 0.75mm Al | |
16 | Additional filtration | 1.00mm Al | |
17 | Half-value layer | ≥1.5mm Al(60kV) | |
18 | Leakage radiation | ≤0.25mGy/h(60kV ,2mA ,2s) | |
19 | Adjustment range of the loading time | 0.01~2.00s,adjustable | |
20 | Distance of the Focal point to skin (The deviation is within 0~5%) | 211mm±3mm | |
21 | X-ray field size at the end of beam limiting tube | round ,Φ 56mm±3mm | |
22 | Nominal electric power | 0.12kW(60kV ,2mA ,0.1s) | |
23 | Available maximum tube current at 60kV | 2mA | |
24 | Inverter frequency of the high voltage generator | 70kHz |
X-ray tube technical specification
Serial numbe | Name | Parameter value |
1 | The X-ray tube model | XD10D-0.5 / 70 |
2 | Nominal value of the focal point | 0.8mm |
3 | Target material | Tungsten |
4 | Maximum working tube voltage | 70kV |
5 | Anode angle of the X-ray tube | 19º |
6 | Intrinsic filtration | 0.75 mm Al |
7 | Filament current | 2.0 A |
8 | Filament voltage | 2.85 ± 0.5 V |
Equipment List
Serial number | Name | Unit | Quantity |
1 | Host | PCS | 1 |
2 | Adapter | PCS | 1 |
3 | Power cable | PCS | 1 |
4 | Hand strap | PCS | |
5 | Neck strap | PCS | 1 |