Customization: | Available |
---|---|
Customized: | Customized |
Certification: | CE, FDA |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Main Body
|
High-strength engineering plastic rack, light, beautiful and corrosion resistant
|
Scope of Application:
|
Adult, Child
|
Gas Source
|
O2: 0.28~0.6MPa, N2O: 0.28~0.6MPa
|
Flow Meter
|
O2: 0.05~1.0L/min,1.1~10L/min
N2O: 0.05~1.0L/min,1.1~10L/min |
O2-N2O Linkage and N2O Stopper
|
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.16±0.02Kpa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply
|
25~75L/min
|
Low Oxygen Pressure Alarm
|
There will be sound alarm when the oxygen pressure < 0.2MPa
|
Vaporizer
|
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator
concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. |
Respiratory Bellows
|
Bellows for adults and children
Tidal volume range:0~1500 ml |
Respiratory Circuit
|
Working mode: All-close, Semi-close, Semi-open
APL: ≤12.5 kPa |
Light
|
Equipped with LED illumination, easy for observation
|
ACGO Port
|
With ACGO port to connect with T circuit or BAIN circuit for open-operation.
|
AGSS Port
|
AGSS exhaust gas collection system can be installed.
|
Display Mode
|
High-definition 10.1"TFT color LCD touch screen display, easy and convenient operation
|
Ventilation Mode
|
IPPV, SIPPV, SIMV, PEEP, SIGH, MANUAL
|
Ventilation Function
|
PEEP,SIGH, IRV
|
Ventilation Parameters
|
Tidal volume:50~1500 ml
Rate:2~99 bpm SIMV rate:2~20 bpm I:E:2:1~1:8 Inspiratory trigger pressure: -10~20 cmHO PEEP:2~20 cmHO Pressure range:5~60 cmHO SIGH:1.5 times the inspiratory time/ 100 times breath |
Parameters For Ventilation Monitoring
|
Tidal Volume, Ventilation Volume, IPPV Rate, SIMV Rate, I:E, Peak Pressure of Airway, Pressure-Time Waveform, Flow Rate-Time
Waveform, PEEP, Inspiratory Trigger Pressure (Average Pressure, FiO2 Oxygen Concentration, Battery Level) |
Security Alarm System
|
Oxygen Concentration Alarm: Upper limit setting range:21%~99%,Low limit setting range:18%~80%
Airway Pressure Alarm: Upper limit setting range:5~60 cmHO,Low limit setting range:0~50cmHO Per-minute Ventilation Volume Alarm: Upper limit setting range:1~20L/min,Low limit setting range: 0~19L/min Tidal Volume alarm: Upper limit setting range:100~1500ml,Low limit setting range 0~1400ml |