Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | ECG |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Items | Specification | ||||||
Lead | Standard 12 leads | ||||||
Acquisition Mode | Simultaneous 12 leads acquisition | ||||||
Measurement Range | ±5mVpp | ||||||
Input circuit | Floating; Protection circuit against Defibrillator effect | ||||||
Input Impedance | ≥50MΩ | ||||||
Input circuit current | <0.05µA | ||||||
Record mode | Automatic: 3CH×4+1R, 3CH×4, 3CHx2+2CHx3,3CHx2+2CHx3+1R,6CHX2; | ||||||
Manual: 3CH, 2CH, 3CH+1R, 2CH+1R; | |||||||
Rhythm: Any lead selectable. | |||||||
Filter | EMG Filter: 25 Hz / 30 Hz / 40Hz/75 Hz / 100 Hz / 150Hz | ||||||
DFT Filter: 0.05 Hz/ 0.15 Hz | |||||||
AC Filter: 50 Hz / 60Hz | |||||||
CMRR | >100dB | ||||||
Patient leak current | <10µA (220V-240V) | ||||||
Frequency Response | 0.05Hz~150Hz (-3dB) | ||||||
Sensitivity | 2.5mm/mV, 5 mm/mV, 10 mm/mV, 20 mm/mV (error:±5%) | ||||||
Anti-baseline Drift | Automatic | ||||||
Time constant | ≥3.2s | ||||||
Noise level | <15µVp-p |
Single package size | 43*20*42 cm |
Single gross weight | 5.5KG |
4 unit per carton | |
package size | 82.5*44.5*45cm |
total gross weight | 24 KG |