Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | Patient Monitor |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Standard Packing List
|
Config. A
|
Config. B
|
Config. C
|
Config. D
|
||
No.
|
Parts No.
|
Description
|
QTY
|
QTY
|
QTY
|
QTY
|
1
|
8310400004
|
FHR Transducer
|
1
|
2
|
1
|
2
|
2
|
8310400003
|
TOCO Transducer
|
1
|
1
|
1
|
1
|
3
|
8310400012
|
Event Marker
|
1
|
1
|
1
|
1
|
4
|
8189007066
|
Belt
|
2
|
3
|
2
|
3
|
5
|
8111810003
|
Recording Paper (112mmx100mmx150p, European Standard)
|
2
|
2
|
2
|
2
|
6
|
8112900002
|
Power Cord
|
1
|
1
|
1
|
1
|
7
|
8510100002
|
User Manual
|
1
|
1
|
1
|
1
|
8
|
8510120001
|
Checking List
|
1
|
1
|
1
|
1
|
9
|
8310100008
|
Ground Cable
|
1
|
1
|
1
|
1
|
10
|
8189007074
|
Adult NIBP cuff with connector, Sac-type (for 25-35cm)
|
0
|
0
|
1
|
1
|
11
|
8189007010
|
NIBP Extension Hose
|
0
|
0
|
1
|
1
|
12
|
8310400014
|
Reusable Adult SpO2 Probe
|
0
|
0
|
1
|
1
|
Technical specifications
Performance Specifications
|
|
Dimension and Weight
|
|
Dimension
|
320mm*260mm*80mm
|
Weight
|
2.8kg(excluding accessories)
|
Power Supply
|
Voltage AC100~240V, 50/60HZ
|
Display
|
- 10.4" color TFT LED (0-135° rotation) - Resolution: 800*600 pixels |
Battery (Built-in)
|
- Type: Rechargeable lithium battery - Charge Cycle: ≥500 times - Working time: 2 hours (optional larger battery for 4-5 hours) |
Recorder (Built-in)
|
- Method: Thermal printer - Paper: Z-fold, thermosensitive - Paper width: 112 mm (4.41 in) - Printing speed: 1/2/3cm/min - Trace: Max. 4 tracks - Recording way: Real-time Recording, Review Printing |
Alarm
|
- Level: Low, medium and high - Indication: Auditory, Visual, Delay of FHR Alarm |
Input Device
|
- Touch screen (option) - Keypad input (Standard) - Mouse/ Keyboard input (option) - USB port for patient data/NST report backup (option) |
System Output & Extensible Interface
|
- Ethernet Network: 1 standard RJ45 socket - USB Port: 1 |
Safety
|
- IEC60601-1 Approved, CE marking according to MDD93/42/EEC - With reference to RoHS Directive 2011/65/EU recasting |
Trend & Reviewing
|
- Cardiotocography (CTG): Waveform Review: 8640 hours - Vital Signs:Trend: 2160 hours
Waveform Review: 2 hours
NIBP Review: 2000 groups
- Alarm Review: 2000 groups |
-Ultrasound (FHR)
|
- Transducer type: 12-Crystal ultrasound transducer - Measure method: Wide beam pulse doppler - FHR Measurement range: 30 to 240 bpm (US Standard) 50 to 210 bpm (Europe Standard) - Accuracy: ±1 bpm - Intensity: <3 mW/cm2 - Ultrasound frequency: 1MHz±3% - Resolution: 1 bpm |
TOCO
|
- Measurement method: Strain gauge sensor element - Measure range: 0 ~ 100 units - Resolution: 1 unit - Nonlinear deviation : ≤±8% - Baseline setting: 0,5,10,15, 20 can be selected - Auto offset correction: 3 second after connecting the transducer - Auto zero adjust: TOCO value is set to zero negative following a negative Measure value for 5 seconds |
Fetal Movement Recording
|
- Automatic fetal movement (AFM): pulsed doppler ultrasound - Automatic fetal movement autograph - Manual fetal movement marker (MFM) |
Maternal SpO2 (Digital Technic)
|
- Measurement Range: 0 ~ 100 % - Resolution: 1 % - Accuracy: ±2% (70% ~ 100%) ±3 % (35% ~ 69%) Unspecified (0 ~ 34%) |
Pulse Rate
|
- Range: 25~250bpm - Accuracy: ±1% or ±1 bpm, whichever is greater - Resolution: 1bpm |
Maternal NIBP (MNIBP)
|
- Method: Oscillometric - Measure mode Manual, Auto, STAT - Measure Interval in AUTO Mode 1~480 min - STAT mode cycle time: Keep 5 minutes, at 5 seconds interval - Measure and Alarm Range: Systolic: 30~270 mmHg Diastolic: 10~220 mmHg Mean: 20~235 mmHg - Static pressure accuracy ±3mmHg - Resolution: 1mmHg - Accuracy: Maximum Mean error ±5mmHg Maximum Standard deviation 8mmHg - Over pressure Protection: Dual protection via software & hardware |
Maternal Temperature
|
|
Technique
|
Thermistor probe (2.25K)
|
-Measuring and Alarm Range
|
0.0 ºC~ 50 ºC (32ºF~ 122ºF)
|
Unit
|
Celsius (ºC), Fahrenheit ( ºF)
|
Resolution
|
0.1ºC
|
Accuracy
|
±0.1ºC
|
NST Report/FHR Interpretation (Option)
|
- Based on Fischer Evaluation Method (monitoring time≥20 Min) - Average Baseline FHR / Number of Acceleration / Number of Deceleration / Number of Small Acceleration / Amplitude Variation /Number of PV - Automatic Score Result: Sore: 8-10 (Excellent) Sore: 5-7 (Suspicious) Score:≤4 ( Fetus Hypoxia) |
Detailed Images