Customization: | Available |
---|---|
Classification: | Biological Diagnostics |
Type: | Biochemistry Analyzer |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Water consumption:
|
≤4L/ hour
|
Operation System:
|
Windows7/8, Windows XP, etc
|
Data storage:
|
Auto save, auto backup
|
Print format:
|
Customized
|
Power supply:
|
A.C 100~ 240V, 50~60Hz, auto adaption
|
Dimension:
|
566mm*559mm*601mm (Length*Width*High)
|
Applies to the hospital within 20 specimens per day.
|
Parameters
|
|
Liver
|
ALT,ALP,GGT,TP,Alb,TBIL,DBIL
|
Heart
|
LDH,CK,CK-MB,AST,a-HBDH
|
Lipids
|
CHO,TG,HDL _C,LDL_ C
|
Lipoprotein
|
APO A1, APO B,Lp(a)
|
Kidney
|
BUN,CRE,UA
|
Lon
|
Na,Mg,Ca,K,P,CI
|
Other
|
CO-CP,AMS,LIP, GLU,IgA,IgG,lgM, C,CRP
|
Type
|
Full automatic/random option/emergency first
|
Test Speed
|
210tests/hr
|
Test method
|
End point, Kinetic, Fixed time, support STAT
|
Cleaning Method
|
With 7 segment 15 steps auto-washing to guaranty cleaning and avoid carryover support Acid and alkali cleaning system and hot
water cleaning |
Wavelength
|
340nm-630nm, customizable
|
Reagent volume
|
20~400ul, step by 1ul
|
Sample volume
|
2~40ul, step by 1ul
|
Reagent tray
|
54 reagent position with onboard cooling system
|
Sample volume
|
45 sample position, support various primary sample tube
|
Reaction tray
|
Direct solid-heating system, 135 reaction cuvettes with auto-washing function, minimum reaction volume 140ul
|
Light source
|
Halogen tungsten lamp. Life is more than 2000 hours
|
Absorbance
|
0~4.0Abs (10mm conversion)
|
Resolution
|
0.0001Abs
|
Calibration
|
Linear (one-point, two-point and multi-point), Logit-log 4P, Spline
|