Hematology analyzer YJ-7200plus
Advanced Platform:
Nucleic acid fluorescence staining integrated with semiconductor laser flow cytometry
Side fluorescent light (SFL)
• Nucleic acid content of cells
Side scatter (SSC)
• Internal cell structure
Forward scatter (FSC)
• Cell volume information
The advanced platform realizes the analysis of the RET channel scatter plot and the high, medium and low fluorescence intensity of the region through the permeation of fluorescence dye into the cell and nucleic acid material binding, to obtained 7 RET parameters (RET%,RET#, IRF, LFR, MFR, HFR, RHE).
|
Once the deviation of PLT-0 in the RET channel and the PLT measured by the sheath impendance method is found to be too large, the software will give a prompt alarm, which can effectively reduce the interference caused by platelet agglu-tination, large platelet, cell debris and other factors on the PLT count.
|
Multiple Test Modes
• Body fluid mode
Supporting body fluid test by proiding reportableparameters(WBC-BF, TC-BF#, MN#, MN%, PMN#, PMN%, RBC-BF)
New Breakthrough
Parameters
|
WBC, NEU%, LYM%, MON%, EDS%, BAS%, NEU#, LYM#, MON#,EOS#, BAS#, IG#, IG%,
RBC, HGB, MCV, HCT, MCH, MCHC, RDW-SD, RDW-CV, NRBC%, NRBC#, PLT, MPV, PCT, POW, P-LCR, P-LCC, RET%,RET#, IRF, LFR, MFR, HFR, RHE, IPF
|
7 body fluid report parameters
|
WBC-BF, TC-BF#, MN#, MN%, PMN#, PMN%, RBC-BF
|
Cyanide-free colorimetry for HGB test
Sheath impedance method for RBC and PLT test
Semiconductor laser flow cytometry for WBC, RBC, NRBC test
|
|
CBC+DIFF
|
120 samples/ hour
|
CBC+DIFF+RET
|
83 samples/ hour
|
BF
|
40 samples/ hour
|
Storage |
At least 500, 000 results
|
Interface
|
RJ45 network interface, USB interface, can be connected to LIS/HIS
|
Size |
835mmx620mmx770mm
|
Weight |
86kg
|