Multifunctional Radiology table |
1 . Table surface movement: electromagnetic lock pedal release. Vertical: ±560m. Horizontal: ± 120mm. 2. Fixed X-ray grid: Line: 40L/cm, Grid Ratio: 10:1, F. D: 100cm. Movement range: ≥ 610 mm. 3 . Radiography specifications: 5" x 7", 8 " x 10", 10" x 12", 11" x 14", 12" x 15", 14" x 14", 14" x 17" 4 . Table surface to film distance: 75 mm. 5 . Table height : 700mm. 6 . Column movement: 180· every 90 positioning, longitudinal stroke ≥1500 mm. 7 . Column movement: 180· every 90 positioning, longitudinal stroke ≥1500 mm. |
|
X ray collimator |
Max irradiation field: SID ( FFD) = 1000mm hours, no less than 430 mm x 430 mm Min irradiation field: 0x0 mm Inherent filter: 1 mm·Al/75 kVp Additional filter: 1 mm· Al and 2 mm· Al Lead leaf control mode: manual Working power: 24 V AC/DC, 150 W ( The lowest safe voltage) |
Picture | Main configuration | Technical Parameters | |
HV-Generator |
Power Supply |
Voltage : 380V/220V± 10% Frequency : 50HZ± 1HZ Resistance: ≤0.3 |
|
Radiography Parameters |
Inverting Frequency : ≥ 30 kHZ Power: ≥50kW Tube Current: 16-500mA, Steps are adjustable Tube Voltage : 45- 125KV, Stepping is 1KV Exposure Time: 0.01-6.3S ,26 steps Selection Mode: 64 kinds of exposure conditions are stored, the operator can choose them freely |
||
APR storage capacity : 240 0 data |