Customization: | Available |
---|---|
Theory: | Autoclave |
Type: | Pressure Steam Sterilization Equipments |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specifications | ||
Model Number | YJ-AN18 | |
Electrical supply | 220-240 VAC | 110-100 VAC |
Chamber (mm, diameter/depth) | Ø230×360 mm | |
Power (VA) | 1100 VAC | |
Sterilization temperature (°C) | 121 °C | 134 °C |
Overall dimensions (mm) | 605 (width)x445(height)x400(depth) | |
Net weight (kg) | 30 kg | |
Sterilization time | 121 °C | 30 min. |
134 °C | 10 min. |
Sterilizer | 1 |
Rack | 1 |
Trays holder | 1 |
Trays | 2 |
Exhaust tube | 2 |
Program | Type | Max. Load (kg) | Max. Load per tray (kg) |
121 °C / 134 °C | Unwrapped solid material | 2.4 | 1.2 |
Code | Description | Proposed solution |
E2 | Chamber overheating | Check temperature sensor connection. |
E4 | Heating overtime | Somewhere is leaking. Check the door seal. Heater malfunction. |
E6 | During operation, the door has been opened or the door micro switch is damaged. | Confirm that the door handle has been closed to the maximum position. Check or replace the door micro switch |
Frequency | Maintenance Operation |
Monthly | Clean the door seal. |
Weekly | Clean the chamber, the trays and the rack. |
Every Year | Replace the door seal. |