Customization: | Available |
---|---|
Certification: | CE, ISO13485 |
Scale: | Small |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
YJ-F4 Patient Monitor (8.0 inches)
Performance Specifications
Technical Parameters
|
||
ECG
|
Lead Mode
|
5 Leads
|
Lead selection
|
I, II, III, avR, avL, avF, V,
|
|
Waveform
|
3 channel
|
|
CMRR
|
Monitor: > 105 dB, Operation: > 105 dB, Diagnosis: > 85 dB
|
|
ECG Signal Range
|
±8 mV (Vp-p)
|
|
Electrode offset Potential
|
± 300 mV
|
|
Leakage Current
|
< 10 uA
|
|
Baseline Recovery
|
< 3 seconds after defibrillation
|
|
Alarm
|
Available
|
|
Review
|
Available
|
|
Respiration
|
Method
|
Impedance between R-F (RA-LL)
|
Differential Input Impedance
|
>2.5 MΩ
|
|
Measurement Impedance Range
|
0.3~5.0 Ω
|
|
Baseline Impedance Range
|
0~2.5 KΩ
|
|
Bandwidth
|
0.3 ~ 2.5 Hz
|
|
Resp. Rate
|
Measurement and Alarm Range: Adult: 0 ~ 120 rpm, Neo/Ped: 0 ~ 150 rpmResolution: 1 rpm Accuracy: ±2 rpm |
|
Apean Alarm
|
10 ~ 40 S
|
|
NIBP
|
Method
|
Oscillometric
|
Mode
|
Manual, Auto, STAT
|
|
Pulse Rate Range
|
40 ~ 240 bpm
|
|
Alarm Type
|
SYS, DIA, MEAN
|
|
Measurement and Alarm Range
|
Adult Mode: SYS: 40 ~ 270 mmHg, DIA: 10 ~ 215 mmHg, MEAN: 20 ~ 235 mmHg Pediatric Mode: SYS: 40 ~ 200 mmHg, DIA: 10 ~ 150 mmHg, MEAN: 20 ~ 165 mmHg Neonatal Mode: SYS: 40 ~ 135 mHg, DIA: 10 ~ 100 mmHg, MEAN: 20 ~ 110 mmHg |
|
Resolution
|
Pressure: 1mmHg
|
|
SpO2
|
Measurement Range
|
0 ~ 100 %
|
Alarm Range
|
0 ~ 100 %
|
|
Resolution
|
1 %
|
|
Accuracy
|
70% ~ 100%: ±2 %, 0% ~ 69%: unspecified
|
|
Actualization interval
|
about 1 second
|
|
Alarm Delay
|
10 second
|
|
Pulse Rate Measurement and Alarm
|
Range: 0 ~ 254 bpm Resolution: 1 bpm Accuracy: ± 2 bpm |
|
Temperature
|
Channel
|
1
|
Measuring and Alarm Range
|
0 ~ 50 °C
|
|
Resolution
|
0.1 °C
|
|
Accuracy
|
± 0.1 °C
|
|
Actualization interval
|
about 1 second
|
|
Average Time Constant
|
10 second
|
Related Products: