| Customization: | Available |
|---|---|
| Type: | Anaesthesia Machine |
| Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency


| Anesthesia Machine | |
| Mode | YJ-A804 |
| Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
| Scope of application | adult, child (4 years above) |
| Gas source | O2, N2O: 0.28~0.6MPa |
| Flow meter | O2, N2O, 0.05-1.0L/min, 1.1-10L/min |
| O2, N2O linkage and N2O Stopped | When using nitrous oxide, oxygen concentration>25%. When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
| Vaporizer | Two, high-precision evaporators |

| Part of Ventilator | |
| Display mode | 5.7" |
| mode | IPPV, SIPPV, SIMV, MANUAL |
| function | PEEP, SIGH, IRV |
| TV | 50~1500ml |
| Rate | 2~99 bpm |
| SIMV rate | 2~20 bpm |
| I/E | 2:1~1:8 |
| Inspiratory trigger pressure | -10~20cmH2O |
| PEEP | 2~20cmH2O |
| Pressure limitation | 10~60cmH2O |
| SIGH | 1.5 times the inspiratory time/100bpm |
| Power | AC 220V 50Hz (UPS, Storage battery) |
| Parameters for ventilation monitoring | B1, B2, B4: tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow - time waveform, PEEP, inspiratory trigger pressure B3, B5: tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, inhale tidal volume, inspiratory platform, average pressure, autonomous inspiratory rate, oxygen concentration, battery capacity, flow-volume loop, pressure-volume loop |
| Mechanical Arms | YES |
| End-tidal CO2monitoring | Optional |