GAS Supply ·Gas source: 3 gas, O2,N2O,Air ·Gas source pressure gauge: 3 gas, O2,N2O,Air ·Output pressure of regulator: 280kpa~300kpa ·Security valve release pressure:<=600 kpa ·Pressure of pipeline inlet: O2,280kPa~600kPa/41Psi~87Psi, <=180L/min N2O,280kPa~600kPa/41Psi~87Psi, <=180L/min Air,280kPa~600kPa/41Psi~87Psi, <=180L/min ·N2O cut off valve protection: when O2 is lost, N2O will be cut off. Flow Meter ·6 tubes high precision flow meter with O2&N2O linkage device ·Gas scale: O2: two tubes, 0~1L/min,1.1L/min~10L/min; N2O: two tubes, 0~1L/min,1.1L/min~12L/min; Air: two tubes, 0~1L/min,1.1L/min~15L/min; ·Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5. ·O2&N2O linkage device, to make insure O2 concentration is not lower than 25% O2 flush: 35L/min~75L/min Select Switch: select between C.G.O and CO2 Circle Absorber C.G..O: alone used for open breath circuit Selectatec Mounting bar: one unit place used for vaporizer, it is easy to replace vaporizer Respiration System ·Absorber chamber:1000mL ·System leakage: when pressure is lower than 3kPa, <=125L/min ·System compliance: <=4ml/kPa ·Resistance of system: Flow rate:30L/min<0.16kPa Flow rate:60L/min<0.52kPa ·APL valve: 0.19~6kPa Dimensions of the machine: ·Height:40cm(without CO2 circle absorber)/68cm (with CO2 circle absorber) ·Width:42cm(without CO2 circle absorber)/66cm (with CO2 circle absorber) ·Depth:20cm(without CO2 circle absorber)/23cm (with CO2 circle absorber) ·Weight: 21kg(with anesthesia vaporizer and without CO2 circle absorber)/ 24kg(with anesthesia vaporizer and with CO2 circle absorber)