Customization: | Available |
---|---|
Classification: | Yj-FC113 Syringe Pump |
Type: | Yj-FC113 Syringe Pump |
Still deciding? Get samples of US$ 340/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Technical specification:
Model | YJ-FC113 |
Applicable syringes | 5ml,10ml,20ml,30ml,50ml/60ml |
Infusion Modes | Rate Mode, Time Mode, Body-Weight Mode |
Drug Name Display | More than 50, Can be updated according to customer requirements or clinical use |
Accuracy | +-2% |
Time Preset | 0-99h and 59min |
Flow Rate step by | Increments at 0.01ml/h |
Flow rate | 5ml: 0.01-150ml/h 10ml: 0.01-400ml/h 20ml: 0.01-600ml/h 30ml: 0.01-900ml/h 50ml: 0.01-1800ml/h 60ml: 0.01-1800ml/h |
Safetly | Double CPU to make sure higher standard infusion |
Volume limit | 0.1-9999.99ml |
Purge/Bolus Rate/Top flow rate | 5ml: 150ml/h 10ml: 400ml/h 20ml: 600ml/h 30ml: 900ml/h 50ml: 1800ml/h 60ml: 1800ml/h |
Audible&visible alarms | syringe dislocation, occlusion , near end, pause overtime, low battery, battery exausted, VTBI completion, syringe empty, motor malfunction, incorrect syringe specification, Circult Malfunction, internal communication malfunction, master CPU malfunction, monitoring CPU malfunction. |
KVO | Rate≥10ml/h: KVO=3ml/h, Rate<10ml/h: KVO=1ml/h |
Occusion pressure | High: 106.7kpa+-26.7kpa, Middle: 66.7Kpa+-13.3Kpa Low: 40.7kpa+-13.3kpa |
Pause overtime alarm | 1min 50s-2min |
Power & Battery | 20VA AC:100V-240V 50HZ/60HZ DC:11.1V Ni-MH battery support from 7 to 10hours |
Calssification | Class II, type CF, IPX4 |
Operation Condition | Temperature: +-5C--+40degree C, Relative Humidity: 20%-90% Atmospheric pressure 70Kpa-106Kpa |
Optional Function | Event log with at least 2000 history records;DC |
Dimensions | 140x260x110mm |
Weight | 2.5kg |