Customization: | Available |
---|---|
Classification: | Imaging Diagnostic Equipment |
Type: | Infusion Pump & Syringe Pump |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Infusion Pump & Syringe Pump 5000 Series
FEATURES OF BMFUSION SERIES
- They can be fixed on an IV stand horizontally and vertically.
- The stackable design allows pumps to be combined freely.
- Numeric keys save operating time.
- Big colorful LCD displays real-time infusion status.
- Three infusion modes: Rate mode, Time mode, and Weight mode, meet different cinical requirements
- Dual chips ensure infusion safety.
- Pipeline pressure is released automatically to reduce infusion dosage.
- Three adjustable occlusion levels meet the different clinical requests
- Emergency Alarm System gives more than 3 minutes alarm tone after power interruption.
- Advanced shock and noise reduction technology keeps the infusion process quiet.
- The clip could be rotated 90°to fix the pumps horizontally and vertically.
- The infusion rate can be changed without pause.
BMFusion 5000A:
|
3 modes (rate mode, time mode, weight mode), 2000pcs history log
|
|||
BMFusion 5000B:
|
2 modes (rate mode, time mode), 2000pcs history log
|
|||
BMFusion 5000C:
|
3 modes (rate mode, time mode, weight mode), 1500pcs history log
|
|||
BMFusion 5000T:
|
4 modes (rate mode, time mode, weight mode, target controlled mode),2000pcs history log
|
Parameter
|
Specification
|
|||
Standard requirements for syringes
|
Compliant with "GB 15810 Single-Use Sterile Syringes"
|
|||
Applicable Syringe Pump Specifications (ml)
|
5ml, 10ml, 20ml, 30ml, 50ml syringes
|
|||
Rate |
5 mL syringe
|
(0. 1~ 150.0)mL/h
|
Step: 0. 1mL/h |
Accuracy: ±2.0% or ±0.005mL/h take the maximum
value |
10 mL syringe
|
(0. 1~300.0)mL/h
|
|||
20 mL syringe
|
(0. 1~600.0)mL/h
|
|||
30 mL syringe
|
(0. 1~900.0)mL/h
|
|||
50 mL syringe
|
(0. 1~ 1500.0)mL/h
|
|||
Bolus rate |
5 mL syringe
|
(0. 1~ 150.0)mL/h
|
Step: 0. 1mL/h |
Accuracy: ±5% |
10 mL syringe
|
(0. 1~300.0)mL/h
|
|||
20 mL syringe
|
(0. 1~600.0)mL/h
|
|||
30 mL syringe
|
(0. 1~900.0)mL/h
|
|||
50 mL syringe
|
(0. 1~ 1500.0)mL/h
|
|||
Occlusion pressure |
20-160KPa
Three levels: high, medium and low. |
The default is medium. |
||
KVO rate
|
Adjustable range: 0. 1~5.0 mL/h (KVO function starts after infusion finished)
|
|||
VTBI range
|
0-9999.9mL
|
|||
Cumulative range
|
0-9999.9mL
|
|||
Single failure bolus
|
About 0.54ml
|
|||
Alarm function |
AC power off, low battery, near complete, complete, occlusion, syringe off, plunger off, reminder, flow error, battery empty, etc
|