Main Functions
|
Hemodialysis; Blood Perfusion (Poisoning Rescue); Simple Ultrafiltration (Dehydration); Liver Ascites Recovery
|
Blood Pump
|
Blood pump flow range: 0.1mL/min~250mL/min
|
Step value: 1mL/min
|
|
Error: ±10%
|
|
Liquid Pump
|
Liquid pump flow range: 2mL/min to 600mL/min
|
Step value: 1mL/min
|
|
Error: ±10%
|
|
Venous Pressure Monitoring
|
Venous pressure range: -13kPa to 66.7kPa (-100mmHg to 800mmHg), resolution: 1mmHg
|
Venous pressure indication accuracy: ±1.3kPa (±10mmHg)
|
|
Adjustable high and low limits for venous pressure alarms, with an alarm action error of ±1.3kPa (±10mmHg)
|
|
Dialysis Hydraulic Monitoring
|
Dialysis hydraulic pressure range: -13kPa to 66.7kPa (-100mmHg to 800mmHg), resolution:1mmHg
|
Dialysis hydraulic pressure indication accuracy: ±1.3kPa (±10mmHg)
|
|
Dialysis hydraulic alarm high limit adjustable, with alarm action error ±1.3kPa (±10mmHg)
|
|
Stable state of continuous operation during 4 hours
|
Fluctuation in dialysate flow rate is of 5%
|
Fluctuation in blood flow rate is of 5%
|
|
Pressure Control
|
The pressure monitoring equipment should have an intravenous pressure and dialysis pressure monitoring system
|
Alarm Action
|
When intravenous pressure monitoring system is alarmed, the equipment shall have the corresponding alarm action: triggering an
audible and visual alarm |
Air Protection System
|
A protection system to prevent safety-related risks to patients due to air entering the body. When air enters the pipeline, the
protection system is activated. An audible and visual alarm is emitted and the blood pump is stopped at the same time, blocking the venous circuit and preventing air bubbles from entering the body. |